STERIM® 10h Biological Ampoule Tests
The test ampoules contain bacterial strains and a nutrient medium. The biologically active material, safely enclosed in the ampoule, is placed in the sterilizer chamber.
After the sterilization cycle, the nutrient medium is added and the bacteria undergo incubation at 55–60°C. Preliminary results can be read after 2.5 hours of incubation, while the final evaluation of bacterial growth capability is performed after 10 hours. The retention of the original color in the biological ampoule confirms microbial inactivation and, therefore, the effectiveness of sterilization.
The packaging of the biological test also includes a chemical indicator confirming that the sterilization process was completed.
Advantages of STERIM® 10h Biological Ampoule Tests:
- Reliable test results, even under stringent sanitary conditions,
- Short incubation period for biological material (preliminary reading after 2.5 hours, final evaluation after 10 hours).
Technical Specifications:
- Sterilization method verified: steam,
- Compliance with standards: ISO 11138, EN 867-3, ANSI/AAMI/ISO 14161:2000,
- Storage: ≤ 27°C,
- Package quantity: 50, 10, or 1 ampoule.
Biological ampoule tests 10h STERIM®
Biological tests with short incubation period. The ampoule contains bacterial spores. After sterilization and a 10-hour incubation, spore survival is assessed.
- Add feedback:
- Manufacturer: Informer Med
Ostatnie sztuki
Najczęściej kupowane z tym produktem
Sprawdź, co jeszcze kupują inni
Flat sterilization roll STERIM®
Sterilization roll with fold STERIM®
Opinie
|
A
|
Anonim ponad pół roku temu
|
|
C
|
Czesław ponad pół roku temu
|
|
e
|
ed ponad pół roku temu
|
Ask a question about the product. Our team will be happy to provide a detailed answer to your inquiry.